cell (potential scan rate = 500 mV/s). The encapsulating solution (1 mL of nanocapsule solution or Tween-80 aqueous solution) is added to the cell, zero time is defined at the point addition is complete, and the solution is stirred for 5 s with a magnetic stir bar. Further cyclic voltammograms are then recorded versus time and the faradaic currents are determined for uptake calculation. The faradaic current at t = 0 s was simply calculated by dividing the faradaic current obtained in the initial cyclic voltammogram (before addition of the encapsulation solution is done) by the dilution factor (6/5 for our experiments). Cyclic voltammograms were not recorded simultaneously with stirring because otherwise the Randles-Sevcik equation used in calculating the aqueous concentration of the probes during uptake would not be valid. The experimental data and the error bars plotted correspond respectively to the average value and to the standard deviation obtained from three different measurements.