142 exhibiting uniformity in Dynaflect deflection measurements. Table 5.2 lists those tests performed on each test pavement section. 5.3.2 Dynaflect Tests Testing with the Dynaflect was accomplished using the standard sensor spacing to identify segments of pavement with fairly uniform deflection response. Each segment was tested at 25-ft. spacings until three or more locations provided essentially identical deflection basins. The modified Dynaflect sensor array was then used to obtain deflection measurements. The sensors were positioned by hand at loca tions designated as 1, 4, 7, and 10, as shown in Figure 4.2. These were the best positions, based on the theoretical study, for separation of layer response. The initial part of the field testing involved placing the extra sensor at position 9 in the modified system (standard position 4). This procedure was later changed to placing one sensor near each Dynaflect loading wheel and the remaining sensors placed at modified positions 4, 7, and 10. In the later case, an average value of Dx was used in the analysis. Appendix A lists Dynaflect deflection measurements from each test section. 5.3.3 Falling Weight Deflectometer Tests The FWD measurements were conducted at approximately the same loca tions as the Dynaflect. In the tests the height of drop and the weight were adjusted to produce different load levels. Two to three drops (load levels) were usually made, with the highest load often repeated. The highest load was generally close to 9 kips. The exact magnitude of each load applied was registered by a load cell located just above the loading plate.